Product Description:

The modular Thermo Capillary Electrophoresis[TM] Crystal CE System joins the Thermo/ATI Unicam Crystal 1000 CE Conductivity Detector with the Thermo 310 Autosampler and HVPS high voltage power supply to provide separations for analyses based on differences in molecular size, shape and charge, as well as entangled polymer solutions and selected stationary phases used in chromatography.


    Crystal 1000 Analyzer (Detector)
  • Detection volume of less than 3 nanoliters permits more sensitive detection of small molecular weight ions than do conventional indirect detection methodologies
  • ConTip[TM] and ConCap[TM] fused silica capillaries are provided with quick finger tight connectors
  • Provides a suite of sensors: Interchangeable ConTip[TM] conductivity sensor; Low Electrolyte Reservoir; Alert/Shutdown Sensor; High-voltage safety interlock sensors
  • Between analyses rinses Auto Detector Cell
  • Includes Temperature Controlled Detector Cell (above ambient to 42 degrees C)
  • Provides semi-automatic cell purge routine for electrolyte conversion
  • Convenient front-panel includes push button electrolyte dispenser, four detector methods memory/recall


  • 2-Lift system Prince Autosampler
  • Dynamic Compression Injection (DCI) generates and applies a range of positive and negative pressures with controlled ramping that lead to reliable and reproducible results; allows control, programming and application of positive and negative injection pressures without the use of external sources and valve even during electromigration of analytes
  • Electrokinetic injection controls ramping in either voltage or current mode in the automated injection system
  • Maintains application environment, eliminating variation in pressure and voltage during the injection process; controls capillary temperature by recycling airflow that is thermostatically regulated by Peltier elements, thus eliminating outside temperature fluctuations
  • High Pressure for CEC and CGE eliminates the need for an external pressure source
  • Carrousel provides accommodation for 30-48 sample/buffer vials and Peltier control of temperature to protect samples from evaporation and degeneration
  • Compatible with other detectors (e.g., UV/VIS, Fluorescence, Diode Array, Mass Spectrometer, and uLIF)


  • PrinCE-HVPS high voltage power supply
  • Enables programmable polarity-switching feature that can either be pre-programmed or activated during injection or run situations, as well as upslope, downslope, and in-run voltages and pressure

Specifications

  • Injection Modes:
  • Hydrodynamic injection: Uses a patented Dynamic Compression Injection (DCI) device to generate and apply a range of positive and negative pressures with controlled ramping, even during electromigration of analytes
  • Electrokinetic injection: features controlled ramping ineither voltage or current mode
  • Injection Pressure: Variable from -180 to +250mbar with 1mbar resolution
  • Flushing/washing: 0 to +2500mbar.
  • Voltage range: 0 to +30kV or -30kV, 100V increments
  • Current range: 0 to +200µA and -200µA, 1µA increments,time-programmable
  • Repeatability: Better than 0.1 % RSD (voltage,current)
  • Ramping: Programmable voltage or current ramping
  • Voltage ramp: 0,1- 50kV/s PrinCE, resolution100V/s, 0,01- 50kV/s PrinCE-C,resolution 10V/s
  • Polarity reversal: Programmable during pre-run and run conditions
  • Autosampler/Fraction Collector: 4-position sample tray or 30/48 Fraction Collector position carrousel Vials randomly accessible from inlet and outlet end of the capillary All systems accept variable vial sizes for samples and buffers,including Eppendorf microcentrifuge tubes
  • Temperature control: Sample/Buffer tray 4 to 40°CCapillary 5°C below ambient to60°C
  • Capillary: Flexible in length and diameter depending on application/configuration
  • Injection duration: Max 650min, resolution 0,01min
  • Repeatability of Injected volume: < 1,0% RSD
  • Power requirements: Mains voltage 115V or 230V, Mains frequency 50/60Hz, Consumption 300VA
  • Dimensions:
  • Height: 50cm (20 inch)
  • Width: 50cm (20 inch)
  • Depth: 54cm (22 inch)
Thermo CE Crystal CE System 310/1000